بيتمنتجاتسلسلة المؤشرات البيولوجيةVHP Space Sterilization Biological Indicator

VHP Space Sterilization Biological Indicator

  • Model Number: TR-VH24
  • Product Description: This product consists of Geobacillus stearothermophilus (ATCC12980) spores and culture medium. Post-incubation color change confirms spore viability, thereby validating the efficacy of VHP sterilization. Spore Load: 1×103 cfu/pc & 1×106 cfu/pc.
  • Intended Use: For verifying the efficacy of vaporized hydrogen peroxide (VHP) space sterilization.

يشارك:

وصف

OPERATING INSTRUCTION

1. Scope of Application

This product is used for verifying the efficacy of vaporized hydrogen peroxide (VHP) space sterilization,

2. Product Specification: TR-VH24

3. Bacterial Strain: Geobacillus stearothermophilus ATCC 12980

4. Population: 1.0×103 CFU/pc & 1.0×106 CFU/pc

5.Instructions for Use
· The biologicalindicator consists of bacterial carriers and culture tubes.
· Placecarriers in different sterilizationzones andexecute theVHP sterilization cycle.
· After sterlization,transfer carriers and culture tubes to a biosafety cabinet.
· Aseptically open carrier packoging and transfer sterlized carriers (plus positive control caier) into culture medium tubes.
· Incubate af 56-58°C for 24 hours.
· Interpret results:
Purple medium: Negative (sterilization effective);
Yellow medium: Positive (sterization ineffective);
Results are valid only if the positive control turns yellow.

  1. Precautions:
    · Store at 10°C to 35°C in dry conditions (30%-70% RH), protected from light (sunlight, fluorescent/UV lamps). Discardafterautoclaving if:Sterlization failureisindicated
    · Past expiration date,Used as positivecontrolVerify product integrity and expiration date before use,
  2. Shelf Life: 24 months.